ENERATION PPB O2 ANALYZER
Xuất sứ: USA
Nhà cung cấp: ANS Vietnam
Hãng sản xuất: Aii1
Ứng dụng sản phẩm: Industrial Products
Liên hệ tư vấn
PI2-UHP-50/100 PPB O2 analyzer, 0-50 PPB low range with LDL <100 PPT for checking purity of ultra-high purity inert gases in the semi-conductor industries, Features
- proprietary Pico-Ion O2 sensor
- LDL less than 0.1 PPB
- 15 months operating sensor life, no periodic maintenance
- integral temperature controlled heated enclosure
- menu driven controls
- fully automated auto-zero, auto-cal
- USB or RS-232 access
- integral data logger
- stainless steel all welded sample system
*Accuracy | +/-3% of reading or 0.25 PPB at constant conditions | Analysis: | 0-50 ppb, 0-1 ppm, 0-10, 0-100 ppm |
Application: | Analysis in ultra-high purity inert gases and H2 | Area Classification: | General purpose |
Alarms: | 2 user adjustable oxygen set-points, “weak sensor”, power failure | Calibration: | Certified 5-8 ppm O2 balance N2 |
Compensation: | Barometric pressure and temperature | LDL: | < 0.1 PPB |
Controls: | Menu driven range selection, calibration, alarm and system functions | Data Acquisition: | Selectable data point intervals |
Display: | Graphical LCD 5 x 2.75; resolution 0.1 ppb | Enclosure: | Painted aluminum 13.9” x 9.9” x 13.4” |
Flow Rate: | 1-2 SCFH recommended | Linearity: | > .995 over all ranges |
Pressure: | Inlet – 20-50 psig; vent – atmospheric | Wetted Parts: | SS 316L |
Recovery Time: | Exposure to 9 ppm for 1 min to less than 10 ppb in 15 minutes | Response Time: | 90% of final FS reading < 60 seconds |
Sample System: | Pressure regulator, Pneumatic valves, sample, span gas inlets, bypass and sensor isolation valves, integral zero scrubber | Connections: | Sample & span inlet – 1/4″ Face Seal, air inlet & sample vent – 1/4″ compression fittings |
Sensor Model: | GPR-13-2000UHP-2/2E | Sensor Life: | 12 months at 25ºC and 1 atm |
Signal Output: | 4-20 mA isolated, 0-5 VDC and 0-1 VDC | Operating Range: | 5º to 45ºC |
Warranty: | 12 months analyzer; 12 months sensor | Power: | 100-240 VAC |
*At constant temperature and pressure
Optional Equipment | |||
Serial Communication | RS-232 | ||
Mounting | 19 rack mount |