GENRAL PURPOSE PPM O2 ANALYZER
Xuất sứ: USA
Nhà cung cấp: ANS Vietnam
Hãng sản xuất: Aii1
Ứng dụng sản phẩm: Industrial Products
Liên hệ tư vấn
GPR-1600 PPM O2 analyzer 0-10 PPM low range, 0.05 PPM sensitivity, general purpose on-line analyzer that measures oxygen concentrations from low PPM to 1%.
Features
- an advanced galvanic oxygen sensor
- over 24 month operating life with no periodic maintenance
- menu driven controls
- integral 4-way Sample/Bypass valve, flow control/meter
- ambient temperature compensation
- two user field selectable oxygen alarms
- USB or RS232 data communication
- integral data logger
utilizing our advanced electrochemical oxygen sensor over a 24 month operating sensor life with no maintenance required, menu driven controls, and a stainless steel integral sample system with a 4-way Sample/Bypass valve that enables isolation of the sensor; a feature that brings the analyzer online and to a low PPM levels in minutes.
Accuracy: | +/-2% of FS range under constant conditions | Analysis Ranges: | 0-10 PPM, 0-100 PPM, 0-1000 PPM, 0-1%, 0-25% FS |
Applications: | Oxygen analysis in inert, hydrogen, mixed and CO2 gas streams | Approvals: | CE |
Area Classification: | General purpose | Alarms: | 2 adjustable relay contacts non-latching; “weak sensor” indicator; power fail; system fail |
Calibration: | Certified gas of O2 balance N2 | Compensation: | Barometric pressure and temperature |
Connections: | 1/4″ compression tube fittings | Controls: | Menu driven range selection, calibration, alarm and system functions |
Data Acquisition: | Selectable data point intervals | Display: | Graphical LCD 5 x 2.75; resolution .01 PPM |
Enclosure: | Painted aluminum 7.5” x 10.8” x 12.25” panel mount | Flow Sensitivity: | None between 1-5 SCFH, 2 SCFH recommended |
Linearity: | > .995 over all ranges | Pressure: | Inlet – regulate to 5-30 psig, max 150 psig; vent – atmosphere |
Power: | 110 or 220 VAC, specify with optional heater | Range ID: | 1-5V output |
Recovery Time: | 60 seconds in air to < 10 ppm in < 1 hr on N2purge | Response Time: | 90% of final FS reading < 10 seconds |
Sample System: | Wetted parts: 316 SS, flow control, 4-way sample/bypass valves and flow indicator | Sensitivity: | < 0.05PPM |
Sensor Model: | GPR-12-333 – requires no maintenance | Sensor Life: | 24 months at 25ºC and 1 atm; average O2< 1000 PPM |
Signal Output: | 4-20mA isolated and 0-1VDC | Temp. Range: | -10ºC to 45ºC |
Warranty: | 12 months analyzer; 12 months sensor |
*At constant temperature and pressure
Optional Equipment | |||
XLT-12-333 | Sensor for gas streams containing CO2 >0.5% | ||
Heater System | PID Temperature Controlled Heating System for applications with varying ambient temperature conditions | ||
Sample System | Coalescing filter, pressure regulator, sample pump |