GPR-2000M PORTABLE O2 ANALYZER GENERAL PURPOSE
Xuất sứ: USA
Nhà cung cấp: ANS Vietnam
Hãng sản xuất: Aii1
Ứng dụng sản phẩm: Industrial Products
Liên hệ tư vấn
Percent oxygen analyzer measures O2 concentrations from 0.05% to 100%. Features an advanced galvanic type oxygen sensor, a 32 month operating life with no periodic maintenance required, user friendly digital electronics, a compact robust enclosure selected for the demands of industrial field use, and a variety of optional equipment.
Accuracy: | < 2% of FS range under constant conditions | Analysis Ranges: | 0-1%, 0-5%, 0-10%, 0-25% (CAL) FS Auto-ranging or manual lock on a single range |
Application: | Oxygen analysis in inert, helium, hydrogen, mixed and acid (CO2) gas streams |
Approvals: | Certified for use hazardous areas – see lower right |
Area Classification: | Class I, Division 1, Groups C, D | Calibration: | 3 month interval using air or certified span gas of O2 approximating 80% of full scale range balance N2 |
Compensation: | Barometric pressure and temperature | Connections: | 1/8″ compression tube fittings |
Controls: | Water resistant keypad; menu driven range selection, calibration and system functions |
Display: | Graphical LCD 2.75 x 1.375”; resolution .001% |
Enclosure: | Painted aluminum, 4 x 9 x 3″, 10 lbs. | Flow: | Not flow sensitive; recommended flow rate 2 SCFH |
LED Indicators: | LOW BATT (72 hr. warning); CHARGE mode | Linearity: | > .995 over all ranges |
Pressure: | Inlet – regulate to 5-30 psig to deliver 2 SCFH flow; vent – atmospheric |
Power: | Rechargeable battery, 60 day cycle, 8 hrs with pump |
Response Time: | 90% of final FS reading in 10 seconds | Sample System: | None; optional integral sample pump and panel mounted flow meter and/or coalescing filter |
Sensitivity: | < 0.5% of FS range | Sensor Model: | GPR-11-32-RTS for non-acid (CO2) gas streams XLT-11-24-RTS for gas mixture with > 0.5% CO2 |
Sensor Life: | GPR-11-32-RTS 32 months in air at 25ºC and 1 atm XLT-11-24-RTS 24 months in air at 25ºC and 1 atm |
Signal Output: | 0-1V FS |
Temp. Range: | 5º to 45ºC (GPR sensor), -20ºto 45ºC (XLT sensor) | Warranty: | 12 months analyzer; 12 months sensor |
Wetted Parts: | Stainless steel flowhousing, tubing and fittings |
Optional Equipment | |||
XLT-11-24 RTS | Sensor for gas mixtures containing CO2 | ||
Filter | Coalescing filter | ||
Flow meter | Sample flow control |