gpr-2500-ais-oxygen-transmitter.png

ATEX and cUL certified analyzer measures O2 concentrations from 0.05% to 100%. Features certification to ATEX Directive 94/9/EC and cUL designed exclusively for Natural Gas applications. This analyzer features an advanced galvanic oxygen sensor, a 36 month operating life, menu driven controls, two (2) field selectable alarm set points, and a stainless steel sample system. Pictured with the optional A-3393 modular sample system designed to remove moderate amounts of free liquids and H2S from natural gas streams

 Accuracy:  < 2% of FS range under constant conditions  Analysis:  0-10 PPM, 0-100 PPM, 0-1000 PPM, 0-1%, 0-25% FS ranges;
auto-ranging or manually lock on single range
 Application:  Oxygen analysis in inert, hydrocarbon, helium, hydrogen,
mixed and acid (CO2) gas streams
 Approvals:  EC TYPE EXAMINATION CERTIFICATE: INERIS 08ATEX0036
II 2 G
Ex d [ib] ib IIB T4
Tamb-20⁰C to +50⁰C
 Area Classification:  Meets the recognized intrinsic safety standards for use in
Class 1, Division 1, Group C, D hazardous
 Alarms:  Two user configurable alarms: magnetic coil relays rated 3A
at 100 VAC, field programmable alarm time delays, alarm
bypass for calibration and system fail alarm
 Calibration:  Air or certified span gas of O2balance N2approximating 80%
of range above analysis range recommended
 Compensation:  Barometric pressure and temperature
 Connections:  1/8″ compression tube fittings  Controls:  Water resistant keypad; menu driven range selection, calibration and system functions
 Display:  Graphical LCD 2.75” x 1.375”; resolution 0.01 PPM; displays
real time ambient temperature and pressure
 Enclosure:  Aluminum ex-proof, fiberglass NEMA 4X, 16x9x6”, 20 lbs.
 Flow Sensitivity:  None between 0.5-5 SCFH, 1-2 SCFH recommended  Linearity:  ±1% of full scale
 Pressure:  Inlet – regulate to 5-30 psig to deliver 1-2 SCFH flow to
transmitter; vent – atmospheric
 Power:  12-24 VDC (suitable for solar power systems) or 100/230 VAC
 Response Time:  90% of final reading in 10 seconds  Sample System:  Sample flow meter
 Sensitivity:  < 0.5% of FS range  Sensor Model:  XLT-12-333 for gases containing > 0.5% CO2
 Sensor Life:  24 months at 25ºC and 1 atm.  Signal Output:  4-20mA
 Operating Range:  Recommended -20ºto 45ºC (XLT sensor)
-10 ºC to 45ºC (GPR sensor)
 Warranty:  12 months analyzer; 12 months sensor
 Wetted Parts:  Stainless steel    
Optional Equipment